142
Appendix D
Error Messages
Error Messages
Error Messages
An error message is displayed in the instrument status display area in the lower-left part of
the screen. Pressing the cursor arrow key on the front panel or executing the :DISP:CCL
command clears the error message. Moreover, about a specific error message, when a
sweep is started again, the display of an error message may disappear. Errors caused by
operation of a front panel key simply appear on the display; with a few exceptions, these
are not stored in the error queue.
The log of an error message, a maximum of 100 pieces are recorded, and it can go back and
check from the oldest error. Executing the :SYST:ERR? command checks error logs. This
operation can be performed only by "SCPI" command. It cannot be operated from the front
panel.
An error with a positive error number is one uniquely defined for this instrument. On the
other hand, an error with a negative error number is basically one defined for common
GPIB devices in IEEE488.2.
1 - 100
21
1 MHz opt. not installed
This error occurs when cable correction commands are sent to E4981A with Option 002,
via GPIB/LAN/USB. This error is not generated with front panel operation.
A
1103
A1 EEPROM write error
An error is generated while writing data to A1 EEPROM.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
B
-168
Block data not allowed
A block data element has been received where the E4981A does not accept any block data
element.
C
41
Correction Measurement Aborted
This error occurs when the correction data measurement is aborted.
1200
CPU bd FLASH ROM write error
An error is generated while writing data to FLASH.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
Содержание E4981A
Страница 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Страница 2: ......
Страница 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Страница 14: ...12 Contents ...
Страница 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Страница 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Страница 96: ...94 Chapter6 Replacement Procedure 3 4 Rack Module Top Replacement Figure 6 3 Screw Fastening Sequence ...
Страница 98: ...96 Chapter6 Replacement Procedure PSU Deck Assembly Removale Figure 6 4 PSU Deck Assembly Removal ...
Страница 100: ...98 Chapter6 Replacement Procedure PSU Assembly Replacement Figure 6 5 PSU Assembly Replacement ...
Страница 102: ...100 Chapter6 Replacement Procedure Front Chassis Assembly Removal Figure 6 6 Front Chassis Assembly Removal ...
Страница 104: ...102 Chapter6 Replacement Procedure Front Panel Assembly Removal Figure 6 7 Front Panel Assembly Removal ...
Страница 108: ...106 Chapter6 Replacement Procedure Interface Board Assembly Replacement Figure 6 10 Screw Fastening Sequence ...
Страница 118: ...116 Chapter6 Replacement Procedure Power Inlet Assembly Replacement Figure 6 17 Power Inlet Assembly Replacement ...
Страница 122: ...120 Chapter6 Replacement Procedure Inverter Assembly Replacement Figure 6 20 Inverter Assembly Replacement ...
Страница 124: ...122 Chapter6 Replacement Procedure Display Support Removal Figure 6 21 Display Support Removal ...
Страница 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Страница 139: ...137 1 Chapter Title 2 Chapter Title 4 Chapter Title 5 Chapter Title C Power Requirement C Power Requirement ...
Страница 142: ...140 AppendixC Power Requirement Power Requirements ...