19-12
Calibration
2.
Select the amps function on Model 6514.
3.
Select Model 6514 20pA range.
4.
Press SHIFT then CAL, then press ENTER at the CAL: RUN prompt. The unit will
prompt for the zero calibration point:
20PA ZERO
5.
Allow the settling time listed in Table 19-5, then press ENTER.
6.
The instrument will prompt for the positive full-scale calibration point:
+20PA CAL
7.
Connect the voltage calibrator and Model 5156 Electrometer Calibration standard to the
Model 6514 INPUT jack, as shown in Figure 19-3. Initially, make connections to the
100G
Ω
resistance.
8.
Press ENTER. The instrument will prompt a full-scale calibration current:
20.00000 PA
9.
Set the calibrator voltage to +2.000000V. Calculate the actual calibration current from
the calibrator voltage and the actual standard resistor value: I = V/R. Adjust Model 6514
display to agree with the calculated current. Allow the settling time listed in Table 19-5,
then press ENTER.
10.
Model 6514 will prompt for the negative full-scale calibration point:
-20PA CAL
11.
Press ENTER. The instrument will prompt for the negative full-scale current:
-20.00000 PA
12.
Set the calibrator output voltage to -2.000000V, then calculate the calibration current
from the calibrator voltage and standard resistor value: I = V/R. Adjust Model 6514 dis-
play to agree with the calculated current. Allow the settling time listed in Table 19-5,
then press ENTER to complete calibration of the present range.
13.
Press EXIT to return to normal display.
14.
Repeat steps 1 through 13 for the 200pA through 2
µ
A ranges using Table 19-5 as a
guide. Be sure to make connections to the correct standard resistor.
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...