![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199326.webp)
Example program
The program below demonstrates how to read back and display calibration standard values.
To use this program, you will need the following:
•
IBM PC compatible computer.
•
Microsoft QBasic (supplied with DOS 5 – 6.22).
•
CEC or Keithley IEEE-488 interface card installed in the computer.
•
Installed CECHP.EXE driver (supplied with above interface card).
' Program to read Model 5156 standard values via Model 6514.
' 6514 primary address = 14.
OPEN
“IEEE” FOR OUTPUT AS #1
' Open IEEE-488 output path.
OPEN
“IEEE” FOR INPUT AS #2
' Open IEEE-488 input path.
#1,
“INTERM CRLF”
' Set input terminator.
#1,
“OUTTERM LF”
' Set output terminator.
#1,
“REMOTE 14”
' Put 6514 in remote.
CLS
“Model 5156 Calibration Standard values:”
#1,
“OUTPUT
14;:CAL:UNPR:KI5156?”
' Query
5156
standard values.
#1,
“ENTER 14”
INPUT
#2,
R100G, R10G, R1G, R100M, C1NF, C100NF
#1,
“LOCAL 14”
CLOSE
“100 G ohm value:”; R100G
“10 G ohm value:”; R10G
“1 G ohm value:”; R1G
“100 M ohm value:”; R100M
“1 nF value:”; C1NF
“100 nF value:”; C100NF
END
Calibration Options
H-3
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...