Volts and Ohms Measurements
3-5
Step 5
Connect the DUT to the electrometer
NOTE
Fundamental information on making connections to the electrometer input is pro-
vided in Section 2 (“Connection Fundamentals”).
WARNING
A metal safety shield is required whenever a hazardous voltage (>30V) is
present on a noise shield or guard shield. As shown in Figures 3-2 and 3-3,
the safety shield must be connected to safety earth ground using #18 AWG
wire or larger.
Unguarded connections
— Connections for unguarded volts and ohms measurements are
shown in Figure 3-2, where the DUT is the voltage or resistance to be measured. If a hazardous
voltage (>30V) is present on the noise shield, or the test circuit is floating above earth ground at
a hazardous voltage level, a safety shield must be used as shown.
RS232
DIGITAL I/O
PREAMP OUT
250V PK
2V ANALOG
OUTPUT
COMMON
CHASSIS
120
FUSE
LINE
630mA
LINE RATING
50, 60Hz
60 VA MAX
T
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
INPUT 250V PK
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
TRIGGER LINK
!
!
MADE IN
U.S.A.
V,
GUARD
(PROGRAMMABLE)
OFF
ON
GUARD
(FOLLOWS
INPUT)
(INTERNAL)
INPUT
PREAMP
10K
PREAMP
OUT
2V ANALOG
OUTPUT
COM
DUT
Black (LO)
Red (HI)
237-ALG-2
Cable
Metal Noise Shield
Metal Safety Shield
6514 Rear Panel
Safety
Earth
Ground
Green
(Chassis)
GRD Disabled
Figure 3-2
Connections for
unguarded volts and
ohms
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...