Digital I/O, Analog Outputs, and External Feedback
11-17
SCPI programming — external feedback
Programming example — external feedback
The following command sequence configures Model 6514 to perform measurements using
the external feedback mode:
SYST:ZCH ON
‘ Enable zero check.
VOLT:XFE ON
‘ Enable external feedback.
FUNC ‘VOLT’
‘ Select Volts function.
SYST:ZCH OFF
‘ Disable zero check.
READ?
‘ Trigger measurement(s) and request
reading(s).
Table 11-4
SCPI commands — external feedback
Commands
Description
Ref
Default
[SENSe]
SENSe Subsystem:
:VOLTage
Volts configuration:
:XFEedback <b>
Enable or disable external feedback.
Off
:FUNCtion VOLTage
Select Volts function.
Sec 3
SYSTem
SYSTem Subsystem:
:ZCHeck <b>
Enable or disable zero check.
Sec 2
On
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...