Triggering
9-7
Counters
Programmable counters are used to repeat operations within the trigger model layers. For
example, if the trigger count is set for 10, operation will keep looping around in the trigger layer
until 10 measurements are performed. If the arm count is set to 2, operation will then loop back
through the arm layer and go back into the trigger layer to perform 10 more measurements.
Trigger model configuration — front panel
NOTE
See “SCPI Programming” (Table 9-2) for the SCPI commands to configure the trig-
ger model.
There are two separate configuration menus to configure the trigger model; one for the arm
layer and one for the trigger layer. Note that trigger delay can be set from either the trigger layer
configuration menu or from the DELAY key.
Once in a menu structure, use the
and
keys to display menu items. Use the cursor (
and
) and the
and
keys to key in values. A menu item or value is selected by pressing ENTER.
Use the EXIT key to exit from the menu.
Arm layer configuration menu
— The configuration menu for the arm layer is structured as
follows. Bullets denote the main items of the menu. To access the menu, press SHIFT and then
CONF-ARM.
•
ARM-IN
— Select the Arm-in control source:
IMM
(Immediate),
GPIB
,
TIMER
(1)
,
MAN
(Manual),
TLINK
(2)
,
/STEST
,
TEST
,
or
BSTEST
.
1.
TIMER — You will be prompted to enter the timer interval (in hour, minute, sec-
ond format). The TIMER annunciator will turn on. The minimum timer setting
is 0.001 seconds.
2.
TLINK — Select the input trigger link line (1 to 6). You will then be prompted
to enable (ONCE) or disable (NEVER) the source bypass.
•
ARM-OUT
— Configure the output trigger:
LINE
— Select the output trigger link line (1 to 6).
EVENT
— Enable (ON) or disable (OFF) the trigger layer done output trigger.
•
COUNT
— Set a finite (FIN) arm count (1 to 2500), or select infinite (INF) arm count.
NOTE
Input trigger and output trigger cannot share the same Trigger Link line. Defaults set
line 1 as the input and line 2 as the output.
Trigger layer configuration menu
— The configuration menu for the trigger layer is struc-
tured as follows. Bullets denote the main items of the menu. To access the menu, press SHIFT
and then CONF-TRIG.
•
TRIG-IN
— Select the Trig-In control source:
IMM
(Immediate), or
TLINK
. After selecting TLINK, select the input trigger link
line (1 to 6). You will then be prompted to enable (ONCE) or disable (NEVER) the
source bypass.
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...