![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199230.webp)
17-10
SCPI Reference Tables
Command
Description
Default
parameter
Ref
SCPI
:ENABle <NDN> or
<NRf>
Program the enable register.
(Note 3)
√
:ENABle?
Read the enable register.
√
:CONDition?
Read the condition register.
√
:QUEStionable
Questionable event registers:
√
[:EVENt]?
Read the event register.
(Note 2)
√
:ENABle <NDN> or
<NRf>
Program the enable register.
(Note 3)
√
:ENABle?
Read the enable register.
√
:CONDition?
Read the condition register.
√
:PRESet
Return status registers to default states.
√
:QUEue
Read error queue:
√
[:NEXT]?
Read and clear oldest error/status (code and
message).
(Note 4)
√
:ENABle <list>
Specify error and status messages for error
queue.
(Note 5)
√
:ENABle?
Read the enabled messages.
√
:DISable <list>
Specify messages not to be placed in queue.
(Note 5)
:DISable?
Read the disabled messages.
:CLEar
Clear messages from error queue.
Parameters:
<NDN> = #Bxx…x
Binary format (each x = 1 or 0)
= #Hx
Hexadecimal format (x = 0 to FFFF)
= #Qx
Octal format (x = 0 to 177777)
<NRf>
= 0 to 65535
Decimal format
<list>
= (100:200, -224)
Example of a range and single entry (+100 t200 and -224)
Notes:
1. Commands in this subsystem are not affected by *RST or SYSTem:PRESet. The effects of cycling power, *CLS and
STATus:PRESet are explained by the following notes.
2. Event registers — Power-up and *CLS clears all bits. STATus:PRESet has no effect.
3. Enable registers — Power-up and STATus:PRESet clears all bits. *CLS has no effect.
4. Error queue — Power-up and *CLS empties the error queue. STATus:PRESet has no effect.
5. Error queue messages — Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.
Table 17-6 (cont.)
STATus command summary
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...