Limit Tests
10-3
A test is only performed if it is enabled. Therefore, you can perform a single-stage test or a
2-stage test. In the flowchart (Figure 10-3), operation simply proceeds through a disabled test.
Table 10-1
Test limit display messages
Display
Message
Limit 1
Test Result
Limit 2
Test Result
:OK
Pass
Pass
:L1
Fail
Not Performed
:L2
Pass
Fail
Pass
?
Start
Measure
DUT
Yes
Limit 1
Test
No
Display
“L1”
Limit 2
Test
Pass
?
Display
“L2”
Display
“OK”
Yes
No
End
Figure 10-3
Operation model
for limit test