![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199222.webp)
17-2
SCPI Reference Tables
General notes
•
Brackets ([ ]) are used to denote optional character sets. These optional characters do not
have to be included in the program message. Do not use brackets in the program
message.
•
Angle brackets (< >) are used to indicate parameter type. Do not use angle brackets in
the program message.
•
The Boolean parameter (<b>) is used to enable or disable an instrument operation. 1 or
ON enables the operation, and 0 or OFF disables the operation.
•
Upper case characters indicated the short-form version for each command word.
•
Default parameter — Listed parameters are both the *RST and SYSTem:PRESet
defaults, unless noted otherwise. Parameter notes are located at the end of each table.
•
Ref — Refers you to the section (Sec) that provides operation information for that com-
mand or command subsystem.
•
SCPI — A checkmark (
√
) indicates that the command and its parameters are SCPI con-
firmed. An unmarked command indicates that it is a SCPI command, but does not con-
form to the SCPI standard set of commands. It is not a recognized command by the SCPI
consortium. SCPI confirmed commands that use one or more non-SCPI parameters are
explained by notes.
Table 17-1
CALCulate command summary
Command
Description
Default
parameter
Ref
SCPI
:CALCulate[1]
Path to configure and control CALC1 calculations.
Sec 7
√
:FORMat <name>
Select math format; MXB or PERCent.
MXB
√
:FORMat?
Query math format.
√
:KMATh
Configure math calculations:
:MMFactor <NRf>
Set “m” for mX+b calculation; -9.99999e20 to
9.99999e20.
1.0
:MMFactor?
Query “m” factor.
:MBFactor <NRf>
Set “b” for mX+b calculation; -9.99999e20 to
9.99999e20.
0.0
:MBFactor?
Query “b” factor.
:MUNits <name>
Specify units for mX+b result; 3 letters (A - Z).
“MXB”
:MUNits?
Query units.
:PERCent
Configure percent math calculation:
[:REFerence] <NRf>
Specify reference value; -9.99999e20 to
9.99999e20.
1.0
:ACQuire
Use input signal as reference value.
:PERCent?
Query reference value.
:STATe <b>
Enable or disable CALC1 calculation.
OFF
√
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...