![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199119.webp)
9-2
Triggering
Trigger models
The flowcharts in Figures 9-1 and 9-2 summarize triggering for Model 6514. They are called
trigger models because they are modeled after the SCPI commands to control triggering
(operation).
Arm-In
Source
Immediate
GPIB
Timer
Manual
TLink
/STest
Test
BSTest
Arm Event
Detector
Trigger Event
Detector
Trigger Delay
MEASURE
Action
No
Yes
Idle
Another
Arm
?
Another
Trigger
?
Yes
No
Arm
Layer
Trigger
Layer
Trigger-In
Source
✛
✛
✛
✛
✛
✛
INF
1
Yes
Halt
?
No
Once
Bypass
Arm Event
Detector
?
Never
Bypass
Trigger Event
Detector
?
Once
✛
Never
Output Trigger
(VMC)
On/Off
Trigger
Count
Output Trigger
(TL Done)
On/Off
Arm
Count
Immediate
TLINk
Press HALT
Turn 6514 ON
✛
0.0 sec
✛
Factory Default
Output Trigger
✛
Figure 9-1
Trigger
model —
front panel
operation
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...