![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199328.webp)
3.
Send the appropriate commands to select the function and range to be calibrated. For ex-
ample, the following commands select the volts function and the 2V range:
:SENS:FUNC ‘VOLT’
:SENS:VOLT:RANG 2
4.
Make appropriate connections, then send the commands for each calibration point for
the selected function and range. For example, send the following commands for the 2V
range:
:CAL:PROT:SENS 0
:CAL:PROT:SENS 2
:CAL:PROT:SENS –2
NOTE
Be sure the appropriate calibration signal is applied to the INPUT jack before send-
ing the command for each calibration point. See Section 19 for details.
5.
Repeat steps 3 and 4 for each function and range.
6.
After all functions and ranges are calibrated, send the commands to program the calibra-
tion dates; for example:
:CAL:PROT:DATE 1998,12,15
:CAL:PROT:NDUE 1999,12,15
7.
Finally, send the following commands to save calibration constants and then lock out
calibration:
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Calibration Options
H-5
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...