Calibration
19-5
Calibration errors
Model 6514 checks for errors after each calibration step, minimizing the possibility that
improper calibration may occur due to operator error. If an error is detected during calibration,
the instrument will display an appropriate error message. The unit will then prompt you to repeat
the calibration step that caused the error.
Calibration menu
You can access the calibration menu by pressing SHIFT then CAL. The various selections are
summarized in Table 19-2. Use the up or down RANGE key to scroll through these selections.
Aborting calibration
You can abort the calibration procedure at any time by pressing the EXIT key.
Table 19-2
Calibration menu
Menu item*
Description
RUN
Calibrates present function and range.
COUNT
Displays calibration count.
IOFFSET
Performs input bias current calibration.
VOFFSET
Performs offset voltage calibration.
DATES
Displays calibration and due dates.
UNLOCK
Unlocks calibration using code.
LOCK
Locks cal, exits to the main menu.
SAVE
Saves calibration constants.
* Press SHIFT then CAL to access. Use up or down RANGE to scroll through selections.
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...