9-8
Triggering
•
TRIG-OUT
— Configure output triggers:
LINE
— Select the output trigger link line (1 to 6).
VMC
— Enable (ON) or disable (OFF) the VMC (voltmeter complete) output trigger.
•
DELAY
— Configure the trigger delay of the trigger layer:
MAN
— Manually set the delay (0 to 999.9998 seconds).
AUTO
— Selects auto delay. The delay is set according to function/range (see Table
9-1).
•
COUNT
— Set a finite (FIN) trigger count (1 to 2500), or select infinite (INF) trigger
count.
NOTE
Input trigger and output trigger cannot share the same trigger link line. Defaults set
line 1 as the input and line 2 as the output.
DELAY Menu
— To access the menu for trigger delay, press the DELAY key.
•
DELAY
— Configure the trigger delay of the trigger layer:
MAN
— Manually set the delay (0 to 999.9998 seconds).
AUTO
— Selects auto delay. The delay is set according to function/range (see Table
9-1).