OHMS
ACCURACY
TEMPERATURE
(1Year)1
COEFFICIENT
TEST
5
1
⁄
2
-DIGIT
18°–28°C
0°–18°C & 28°–50°C
CURRENT
RANGE
RESOLUTION
±(% rdg+counts) ±(% rdg+counts)/°C
(nominal)
2 k
Ω
10 m
Ω
0.20+ 10
0.01 + 2
0.9 mA
20 k
Ω
100 m
Ω
0.15+ 3
0.01 + 1
0.9 mA
200 k
Ω
1
Ω
0.25+ 3
0.01 + 1
0.9 mA
2 M
Ω
10
Ω
0.25+ 4
0.02 + 2
0.9 µA
20 M
Ω
100
Ω
0.25+ 3
0.02 + 1
0.9 µA
200 M
Ω
1 k
Ω
0.30+ 3
0.02 + 1
0.9 µA
2 G
Ω
10 k
Ω
1.5 + 4
0.04 + 2
0.9 nA
20 G
Ω
100 k
Ω
1.5 + 3
0.04 + 1
0.9 nA
200 G
Ω
1 M
Ω
1.5 + 3
0.04 + 1
0.9 nA
1 When properly zeroed, 5
1
⁄
2
digit. Rate: Slow (100ms integration time).
MAXIMUM OPEN CIRCUIT VOLTAGE:
250VDC.
PREAMP SETTLING TIME (To 10% of final reading with <100pF input capacitance):
2k
Ω
through 200k
Ω
: 2ms; 20M
Ω
through 200M
Ω
: 90ms. 2G
Ω
through 200G
Ω
: 1s.
COULOMBS
ACCURACY
TEMPERATURE
(1 Year)1,2
COEFFICIENT
6
1
⁄
2
DIGIT
18°–28°C
0°–18°C & 28°–50°C
RANGE
RESOLUTION
±(%rdg+counts)
±(%rdg+counts)/°C
20 nC
10 fC
0.4 + 50
0.04 + 10
200 nC
100 fC
0.4 + 50
0.04 + 10
2 µC
1 pC
1 + 50
0.05 + 10
20 µC
10 pC
1 + 50
0.05 + 10
Notes:
1 Charge acquisition time must be <1000s, derate 2% for each additional 10,000s.
2 When properly zeroed, 6
1
⁄
2
digit. Rate: Slow (100ms integration time).
INPUT BIAS CURRENT:
<4fA at TCAL. Temperature coefficient = 0.5fA/°C.
IEEE-488 BUS IMPLEMENTATION
MULTILINE COMMANDS:
DCL, LLO, SDC, GET, GTL, UNT, UNL, SPE, SPD.
IMPLEMENTATION:
SCPI (IEEE-488.2, SCPI-1996.0); DDC (IEEE-488.1).
UNILINE COMMANDS:
IFC, REN, EOI, SRQ, ATN.
INTERFACE FUNCTIONS:
SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1.
PROGRAMMABLE PARAMETERS:
Function, Range, Zero Check, Zero Correct, EOI
(DDC mode only), Trigger, Terminator (DDC mode only), Data Storage 2500 Storage,
Calibration (SCPI mode only), Display Format, SRQ, REL, Output Format, Guard,
V-offset Cal, I-offset Cal.
ADDRESS MODES:
TALK ONLY and ADDRESSABLE.
LANGUAGE EMULATION:
6512, 617, 617HIQ emulation via DDC mode.
TRIGGER TO READING DONE:
150ms typical, with external trigger.
RS-232 IMPLEMENTATION:
Supports
: SCPI 1996.0.
Baud Rates:
300, 600, 1200, 2400, 4800, 9600, 19.2k, 38.4k, 57.6k.
Protocols:
Xon/Xoff, 7 or 8 bit ASCII, parity-odd/even/none.
Connector:
DB-9 TXD/RXD/GND.
Specifications
A-3
Section A pg 2-4 5/5/03 1:57 PM Page 2
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...