19-18
Calibration
Entering calibration dates and saving calibration
NOTE
For temporary calibration without saving new calibration constants, proceed to Lock-
ing out calibration.
1.
Press SHIFT then CAL to access the calibration menu.
2.
Use either RANGE key to display the following:
CAL: DATES
3.
Press ENTER. The unit will display:
DATE: 06/15/98
4.
Use the arrow and RANGE keys to set the date, then press ENTER. The unit will then
prompt for the calibration due date:
NDUE: 06/15/99
5.
Set the due date as desired, then press ENTER.
6.
Select CAL: SAVE from the calibration menu, then press ENTER. The unit will prompt
as follows:
SAVE CAL?YES
7.
With the YES prompt displayed, press ENTER to save and lock out calibration. The unit
will display:
CAL SAVED
NOTE
Calibration will also be locked out once saved.
Locking out calibration
Use the following procedure to lock out calibration without saving new calibration constants:
1.
Press SHIFT then CAL, then use the up RANGE key to display the following:
CAL: LOCK
2.
Press ENTER. The instrument will display the following message:
CAL LOCKED.
Changing the calibration code
Follow the steps below to change the calibration code:
1.
Press SHIFT then CAL. The instrument will display the following:
CAL: RUN
2.
Use the up or down RANGE key to display the following:
CAL: UNLOCK
3.
Press ENTER. The instrument will prompt for the present calibration code:
CODE?
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...