17-8
SCPI Reference Tables
Command
Description
Default
parameter
Ref
SCPI
:LGRoup <name>
Specify autorange limit; HIGH or LOW.
HIGH
:LGRoup?
Query upper limit for autorange.
:AUTO?
Query state of autorange.
√
:ADIScharge
Path for auto discharge:
Sec 5
:LEVel <NRf>
Set auto discharge level; -21e-6 to 21e-6.
2e-6
:LEVel?
Query auto discharge level.
[:STATe] <b>
Enable or disable auto discharge.
OFF
[:STATe]?
Query state of auto discharge.
:AVERage
Path to control the Digital Filter:
Sec 6
:TCONtrol <name>
Select filter control; MOVing or REPeat.
REP
:TCONtrol?
Query filter control.
:COUNt <n>
Specify filter count; 2 to 100.
10
:COUNt?
Query filter count.
[:STATe] <b>
Enable or disable digital filter.
OFF
[:STATe]?
Query state of digital filter.
:MEDian
Path to control median filter:
Sec 6
:RANK <NRf>
Specify “n” for rank; 1 to 5 (rank = 2n+1).
1
:RANK?
Query rank.
[:STATe] <b>
Enable or disable median filter.
OFF
[:STATe]?
Query state of median filter.
Note: *RST default is ON and SYSTem:PRESet default is OFF.
Table 17-4 (cont.)
SENSe command summary
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...