![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199142.webp)
Limit Tests
10-11
•
LIN4MOD
(Line 4 Mode):
ENDOFTST
(End of Test) — With this mode, Model 6514 will pulse the EOT line
when the test is finished. Use with catagory register component handlers.
/BUSY
and
BUSY
— Pulls line 4 LO (/Busy) or HI (Busy) while the test is in pro-
cess. Use with catagory pulse component handlers.
Arm layer configuration menu
When using a component handler, the “start of test” option is selected from the Arm-In menu
item of the arm layer configuration menu. To access the menu, press SHIFT and then CONF-
ARM.
•
ARM-IN
— Select the “start of test” option:
IMM
(Immediate) — Test starts when LIMIT key is pressed.
TEST
— Test starts when the handler pulls the SOT line of the Digital I/O high.
/STEST
— Test starts when the handler pulls the SOT line of the Digital I/O low.
NOTE
The other arm-in control sources are seldom used with component handlers, but are
available.
Perform limit tests
Step 1
Configure test system
As previously explained, testing the system could be as simple as connecting a DUT to Model
6514 or could employ the use of a component handler for binning operations.
Step 2
Configure measurement
Configure Model 6514 for the desired measurement as covered in the previous sections of this
manual.
Step 3
Configure limit tests
Configure Model 6514 for the limit tests as explained in “Limit Test Configuration”.
Step 4
Start testing process
To enable the limit tests, press the LIMIT key. If the SOT line of the Digital I/O port is being
used by a component handler, the test process will not start until the handler pulses the SOT line.
Otherwise, the testing process will start when LIMIT is pressed.
Step 5
Stop testing process
The testing process can be terminated at any time by again pressing the LIMIT key. When
using a component handler, the testing process will stop after the last DUT is tested.
Содержание 6514
Страница 203: ...14 CommonCommands...
Страница 207: ...15 SCPISignalOriented MeasurementCommands...
Страница 235: ...18 PerformanceVerification...
Страница 254: ...19 Calibration...
Страница 274: ...20 RoutineMaintenance...
Страница 278: ...A Specifications...
Страница 282: ...B StatusandErrorMessages...
Страница 288: ...C GeneralMeasurement Considerations...
Страница 294: ...D DDCEmulationCommands...
Страница 299: ...E ExamplePrograms...
Страница 306: ...F IEEE 488BusOverview...
Страница 320: ...G IEEE 488andSCPI ConformanceInformation...
Страница 324: ...H CalibrationOptions...