Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-336
3700AS-901-01 Rev. D/June 2018
scan.trigger.sequence.set()
This function sets the sequence even detector to the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
scan.trigger.sequence.set()
Details
This function sets the sequence event detector to the detected state.
Example
scan.trigger.sequence.set()
Sets the sequence event detector to the
detected state.
Also see
(on page 11-335)
(on page 5-23)
scan.trigger.sequence.stimulus
This attribute selects the trigger stimulus for the sequence event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Instrument
reset
Recall setup
Scan reset
Create configuration script
Save setup
28
(scan.trigger.EVENT_CHANNEL_READY)
Usage
eventID
= scan.trigger.sequence.stimulus
scan.trigger.sequence.stimulus =
eventID
eventID
The trigger stimulus that is used for the sequence event
Details
This attribute selects the events that cause the sequence event detector to enter the detected state.
Use this to start a set of measurement count readings that are triggered by a single event.
To pace each reading by an event, use
scan.trigger.measure.stimulus
.
To bypass pacing the readings, set this to 0.