Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-403
Usage
enableRegister
= status.system5.condition
enableRegister
= status.system5.enable
enableRegister
= status.system5.event
enableRegister
= status.system5.ntr
enableRegister
= status.system5.ptr
status.system5.enable =
enableRegister
status.system5.ntr =
enableRegister
status.system5.ptr =
enableRegister
enableRegister
The status of the system summary 5 register; a zero (0) indicates no bits set; other
values indicate various bit settings
Details
In an expanded system (TSP-Link), these attributes are used to read or write to the system summary
registers. They are set using a constant or a numeric value, but are returned as a numeric value. The
binary equivalent of the value indicates which register bits are set. In the binary equivalent, the least
significant bit is bit B0, and the most significant bit is bit B15. For example, if a value of
1.02
(which is 130) is read as the value of the condition register, the binary equivalent is
0000 0000 1000 0010. This value indicates that bit B1 and bit B7 are set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
0
1
0
0
0
0
0
1
0
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page C-19). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable
B1
status.system5.NODE57
Bit B1 decimal value: 2
B2
status.system5.NODE58
Bit B2 decimal value: 4
B3
status.system5.NODE59
Bit B3 decimal value: 8
B4
status.system5.NODE60
Bit B4 decimal value: 16
B5
status.system5.NODE61
Bit B5 decimal value: 32
B6
status.system5.NODE62
Bit B6 decimal value: 64
B7
status.system5.NODE63
Bit B7 decimal value: 128
B8
status.system5.NODE64
Bit B8 decimal value: 256
B9-B15
Not used
Not applicable