Section 6: Reading buffers
Series 3700A System Switch/Multimeter Reference Manual
6-6
3700AS-901-01 Rev. D/June 2018
Buffer configuration (front panel)
When configuring the buffer through the front panel, you must select a reading buffer. See
(on page 6-3) for more information.
To configure a buffer from the front panel:
1. Press the
CONFIG
key.
2. Press the
STORE
key. The RD BUFFER ATTR menu opens.
3. To view the count and capacity of a selected buffer, select the
COUNT
or
CAPACITY
menu
choice. To configure the buffer's append mode, select
APPEND
, then
ON
or
OFF
.
Appending readings
When the buffer append mode is disabled, the buffer is cleared (readings are lost) before a new
storage operation starts.
When buffer append mode is enabled, the buffer is not cleared and each subsequent storage
operation appends the readings to the buffer. When the buffer is filled to capacity, the storage
process stops. The readings must be cleared before the next storage operation starts.
(on page 11-16) for more information.
Buffers created on the front panel have the append mode enabled by default.
Remote buffer operation
You can control the Series 3700A buffer using remote commands.
Data store (buffer) commands
The following commands are associated with data store operation:
•
(on page 11-146)
•
(on page 11-152)
•
(on page 11-153)
•
(on page 11-154)
•
(on page 11-154)
•
(on page 11-196)
•