Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-408
3700AS-901-01 Rev. D/June 2018
Details
Indicates if an event was ignored because the event detector was already in the detected state when
the event occurred. This is an indication of the state of the event detector that is built into the event
blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any
other trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output is
true
.
If an event was not ignored, the output is
false
.
Also see
(on page 11-408)
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.blender[
N
].reset()
N
The trigger event blender (up to two)
Details
The
trigger.blender[
N
].reset()
function resets the following attributes to their factory
defaults:
•
trigger.blender[
N
].orenable
•
trigger.blender[
N
].stimulus[
M
]
It also clears
trigger.blender[
N
].overrun
.
Example
trigger.blender[1].reset()
Resets the trigger blender 1 settings to
factory defaults.
Also see
(on page 11-407)
(on page 11-407)
trigger.blender[N].stimulus[M]