trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Trigger blender N reset
Recall setup
Create configuration script
Save setup
false (AND mode)
Usage
orenable
= trigger.blender[
N
].orenable
trigger.blender[
N
].orenable =
orenable
orenable
The type of operation:
▪
true
: OR operation
▪
false
: AND operation
N
The blender number (up to two)
Details
This command selects whether the blender waits for any one event (
OR
) or waits for all selected
events (
AND
) before signaling an output event.
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
Generate a trigger blender 1
event when a digital I/O
trigger happens on line 3 or
5.
Also see
(on page 11-408)
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Instrument reset
Trigger blender
N
clear
Trigger blender
N
reset
Not applicable
Not applicable
Usage
overrun
= trigger.blender[
N
].overrun
overrun
Trigger blender overrun state (
true
or
false
)
N
The blender number (up to two)