Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-43
6. Send reading.
7. Set voltage,
–12 V to +12 V range, generate positive point 2.
8. Send reading.
9. Set current, 0 mA to +20 mA range, generate point 1.
10. Send reading.
11. Set current, 0 mA to +20 mA range, generate point 2.
12. Send reading.
13. Set current, +4 mA to +20 mA range, generate point 1.
14. Send reading.
15. Set current, +4 mA to +20 mA range, generate point 2.
16. Send reading.
For totalizer channels, a calibration sequence includes these steps:
1. Calibrate 0 V totalizer threshold
2. Calibrate 1.5 V totalizer threshold
You must save the calibration after calibrating and before locking. Use
channel.calibration.save()
to save the calibration.
All calibration progress is lost if the calibration data is not saved before you lock the channel.
After calibration, the channel must be locked using
channel.calibration.lock()
.
Also see
(on page 11-40)
(on page 11-41)
(on page 11-44)