Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-171
dmm.detectorbandwidth
This attribute sets the AC detector bandwidth setting for the DMM in Hertz.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
DMM reset
Recall setup
Create configuration script
Save setup
300
Usage
value
= dmm.detectorbandwidth
dmm.detectorbandwidth =
value
value
The detector bandwidth in Hertz
Details
Only applies when
dmm.func
is set to
"acvolts"
or
"accurrent"
.
If you query this value for any other function, nil is returned.
When you send this value, the input value is adjusted as follows:
Write value
Read value
< 30
3
Between 30 and 300
30
300
300
An error is generated:
•
If you send this command for any function other than
"accurrent"
or
"acvolts"
•
If you set
dmm.aperture
and the detector bandwidth read value is 30 or less
•
If the value is below 3
Example
reset()
dmm.func = "acvolts"
print(dmm.func, dmm.detectorbandwidth)
dmm.detectorbandwidth = 35
print(dmm.func, dmm.detectorbandwidth)
dmm.func = "accurrent"
print(dmm.func, dmm.detectorbandwidth)
dmm.func = "acvolts"
print(dmm.func, dmm.detectorbandwidth)
Sets the detector bandwidth to 35 Hz for
AC volts. 35 is adjusted to 30. AC current is
still at 300 Hz.
Output:
acvolts 3.00002
acvolts 3.00001
accurrent 3.00002
acvolts 3.00001
Also see
(on page 11-145)
(on page 11-166)