Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-327
scan.stepcount
This attribute returns the number of steps in the present scan.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
scanStepCount
= scan.stepcount
scanStepCount
The present step count value
Details
This is set by the number of steps in the active scan list. The value of this attribute is initially
determined when the scan is created. Adding steps with the
scan.create(),
scan.addimagestep()
, and
scan.add()
functions updates this attribute's value.
Example
print(scan.stepcount)
Responds with the present step count.
Output assuming there are five steps in the
scan list:
5.00
Also see
(on page 11-309)
(on page 11-311)
(on page 11-315)
(on page 5-23)
scan.trigger.arm.clear()
This function clears the arm event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
scan.trigger.arm.clear()
Details
This function sets the trigger model's arm event detector to the undetected state.
Example
scan.trigger.arm.clear()
Clears the arm event detector.
Also see
(on page 11-328)
(on page 5-23)