Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-328
3700AS-901-01 Rev. D/June 2018
scan.trigger.arm.set()
This function sets the arm event detector to the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
scan.trigger.arm.set()
Details
This function sets the arm event detector of the trigger model to the detected state.
Example
scan.trigger.arm.set()
Sets the arm event detector to the detected
state.
Also see
(on page 11-327)
(on page 11-328)
(on page 5-23)
scan.trigger.arm.stimulus
This attribute determines which event starts the scan.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Scan reset
Create configuration script
Save setup
0
Usage
eventID
= scan.trigger.arm.stimulus
scan.trigger.arm.stimulus =
eventID
eventID
Trigger stimulus used for the channel action (arm layer); see
Details
Details
This attribute selects which events cause the arm event detector to enter the detected state.
Set this attribute to 0 to start the scan without waiting for an event.