Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-235
errorqueue.clear()
This function clears all entries out of the error queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
errorqueue.clear()
Details
See the
Error Available Bit (error or event queue)
(on page C-8) topic for additional information about
the error queue.
Also see
(on page 11-235)
errorqueue.count
This attribute gets the number of entries in the error queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Power cycle
Clearing error queue
Reading error messages
Not applicable
Not applicable
Usage
count
= errorqueue.count
count
The number of entries in the error queue
Example
count = errorqueue.count
print(count)
Returns the number of entries in the error
queue.
The output below indicates that there are
four entries in the error queue:
4.00
Also see
Error Available Bit (Error or Event Queue)
(on page C-8)
(on page 11-235)