Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-416
3700AS-901-01 Rev. D/June 2018
Example
print(trigger.timer[1].overrun)
If an event was ignored, the output is
true
.
If the event was not ignored, the output is
false
.
Also see
(on page 11-416)
trigger.timer[N].passthrough
This attribute enables or disables the timer trigger pass-through mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Create
configuration script
Save setup
false (disabled)
Usage
passthrough
= trigger.timer[
N
].passthrough
trigger.timer[
N
].passthrough =
passthrough
passthrough
The state of pass-through mode; set to one of the following values:
▪
true
: Enabled
▪
false
: Disabled
N
Trigger timer number (1 to 4)
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
(on page 11-416)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.timer[
N
].reset()
N
Trigger timer number (1 to 4)