Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-168
3700AS-901-01 Rev. D/June 2018
Example
dmm.func = "dcvolts"
dmm.reset("active")
dmm.nplc = 0.5
dmm.range = 10
dmm.configure.set("TestDcv")
dmm.configure.recall("dcvolts")
print(dmm.func, dmm.autorange, dmm.range, dmm.nplc)
dmm.configure.recall("TestDcv")
print(dmm.func, dmm.autorange, dmm.range, dmm.nplc)
Set the DMM to the DC volts function.
Reset DC volts back to factory defaults.
Set the NPLC for DC volts to 0.5.
Select the 10 V range for DC volts and disable autorange.
Save a user DMM configuration for DC volts called "TestDcv".
Recall and configure the DMM for factory DC volts.
Output the settings for factory DC volts.
Recall the user DMM configuration called "TestDcv".
Output the settings for the TestDcv configuration.
Output:
dcvolts
1.00000
1.00001
1.00000
dcvolts
0.00000
1.00001
5.000000000e-01
Also see
(on page 11-110)
(on page 11-161)
(on page 11-162)
(on page 11-163)
(on page 11-179)
(on page 11-227)
dmm.connect
This attribute indicates how the DMM relays should be connected to the analog backplane.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
DMM reset
Recall setup
Create configuration script
Save setup
7 (dmm.CONNECT_ALL)
Usage
value
= dmm.connect
dmm.connect =
value
value
The DMM relay connection setting. See
Details
for valid values