INSTRUCTION MANUAL
KEW MATE 2000/2001
DIGITAL MULTIMETER
WITH
AC/DC CLAMP SENSOR
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.
1.SAFETY WARNINGS
This instrument has been designed and tested according to IEC Publication 61010:
Safety Requirements for Electronic Measuring Apparatus. This instruction manual
contains warnings and safety rules which must be observed by the user to ensure
safe operation of the instrument and to retain it in safe condition. Therefore, read
through these operating instructions before starting using the instrument.
WARNING
● Read through and understand instructions contained in this manual before
starting using the instrument.
● Save and keep the manual handy to enable quick reference whenever
necessary.
● Be sure to use the instrument only in its intended applications and to follow
measurement procedures described in the manual.
● Be sure to understand and follow all safety instructions contained in the manual.
Failure to follow the above instructions may cause injury, damage to the instrument
and/or damage to equipment under test.
The symbol indicated on the instrument means that the user must refer to related
parts of the manual for safe operation of the instrument. Be sure to carefully read the
instructions following each symbol in this manual.
DANGER
is reserved for conditions and actions that are likely to cause serious or fatal
injury.
WARNING
is reserved for conditions and actions that can cause serious or fatal injury.
CAUTION
is reserved for conditions and actions that can cause minor injury or
instrument damage.
Following symbols are used on the instrument and in the instruction manual. Attention
should be paid to each symbol to ensure your safety.
Refer to the instructions in the manual.
This symbol is marked where the user must refer to the instruction manual so as not
to cause personal injury or instrument damage.
Indicates an instrument with double or reinforced insulation.
Indicates that this instrument can clamp on bare conductors when measuring
a voltage corresponding to the applicable Measurement category, which is marked
next to this symbol.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates AC and DC.
DANGER
● Never make measurement on circuits with a maximum voltage difference of
600VAC/DC or greater between conductors (300VAC/DC or greater between a
conductor and ground).
● Do not attempt to make measurement in the presence of flammable gasses.
Otherwise, the use of the instrument may cause sparking, which leads to an
explosion.
● Never attempt to use the instrument if its surface or your hand is wet.
●Do not exceed the maximum allowable input of measuring ranges.
● Never open the battery compartment cover while making measurement.
● Never try to make measurement if any abnormal conditions, such as broken
Transformer jaws or case is noted.
● The instrument is to be used only in its intended applications or conditions.
Otherwise, safety functions equipped with the instrument doesn t work, and
instrument damage or serious personal injury may be caused.
WARNING
● Never attempt to make any measurement, if any abnormal conditions are noted,
such as broken case, cracked test leads and exposed metal parts.
● Do not turn the Function Selector Switch while the test leads are connected to
the circuit under test.
● Do not install substitute parts or make any modification to the instrument. Return
the instrument to Kyoritsu or your distributor for repair or re-calibration.
● Do not try to replace the batteries if the surface of the instrument is wet.
● Always disconnect the clamp sensor and the test leads from the circuit under test
and switch off the instrument before opening the battery compartment cover for
battery replacement.
CAUTION
● Make sure that the Function Selector Switch is set to an appropriate position
before making measurement.
● Always make sure to place the test leads in the test lead holder before making
current measurement.
● Do not expose the instrument to the direct sun, extreme temperatures or dew fall.
● Be sure to set the Function Selector Switch to the "OFF" position after use.
When the instrument will not be used for a long period of time, place it in storage
after removing the batteries.
● Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
Measurement Category:
To ensure safe operation of measuring instruments, IEC 61010 establishes safety
standards for various electrical environments, categorized as O to CAT.IV, and called
measurement categories. Higher-numbered categories correspond to electrical
environments with greater momentary energy, so a measuring instrument designed
for CAT.III environments can endure greater momentary energy than one designed
for CAT.II.
O
: Circuits which are not directly connected to the mains power supply.
CAT.II : Electrical circuits of equipment connected to an AC electrical outlet by a
power cord.
CAT.III : Primary electrical circuits of the equipment connected directly to the
distribution panel, and feeders from the distribution panel to outlets.
CAT.IV : The circuit from the service drop to the service entrance, and to the power
meter and primary over-current protection device (distribution panel).
2.FEATURES
● Permits AC/DC current measurement up to 60A using a clamp sensor that comes
standard with the instrument
●Clamp sensor for ease of use in crowded cable areas and other tight places
● Permits current measurement with an open current-clamp sensor that does not require
opening and closing operations by the user
●Auto-power-save function
●Buzzer for easy continuity checking
●Data hold function to freeze the readings
●LCD with a 3400 count full scale bar graph
●Shock absorbing holster for ease of storage
● Designed to international safety standard IEC61010-1: measurement category
CAT.
Ⅲ
, 300V and pollution degree 2.
O: Device which is
not directly
connected to the
mains power supply
3.SPECIFICATIONS
●Measuring Ranges and Accuracy (at 23℃±5℃, relative humidity75% or less)
※Electromagnetic RF field ≦ 1 V/m
compatibility ACV/DCV/OHMS/FREQUENCY total accuracy = specified accuracy
(IEC 61000-4-3) ACA/DCA total accuracy = specified a5dgt
RF transmitters such as mobile telephones may not be used in close
proximity.
●Safety Standard
IEC 61010-1
measurement CAT.
Ⅲ
, 300V, pollution degree 2
measurement CAT.
Ⅱ
, 600V, pollution degree 2
IEC 61010-031
IEC 61010-2-032, IEC 61010-2-033
IEC 61326-1 (EMC), EN 50581 (RoHS)
●Operating System
Dual integration
●Display
Liquid crystal display with maximum reading of 3399 as
well as units and annunciators
Bar graph with maximum points of 33
●Over Input Indication
"OL" on the LCD (Ω ranges only)
●Auto-ranging Operation
Shifts to the next higher range when bar graph increases
to 33 points
Shifts to the next lower range when bar graph decreases
to 3 points
●Sample Rate Numeric reading: about 400ms,
bar graph: about 20ms
MODEL
2000
2001
Range
60A
100A
AC Current A
DC Current A
AC Voltage V Input impedance: 10MΩ
DC Voltage V Input impedance: 10MΩ
Measuring Range
0-60.0A
0-100.0A
Accuracy
± 2.0%rdg ± 5dgt(50/60Hz)
± 2.0%rdg ± 5dgt(50/60Hz)
MODEL
2000
2001
Range
60A
100A
Measuring Range
0- ± 60.0A
0- ± 100.0A
Accuracy
± 2.0% rdg ± 5dgt
± 2.0% rdg ± 5dgt
Range
3.4V
34V
340V
600V
Measuring Range
0-600V
(Auto-ranging)
Accuracy
± 1.5%rdg ± 5dgt(50-400Hz)
Range
340mV
3.4V
34V
340V
600V
Measuring Range
0- ± 600V
(Auto-ranging)
Accuracy
± 1.5%rdg ± 4dgt
Resistance Ω/
Range
340 Ω
3.4k Ω
34k Ω
340k Ω
3.4M Ω
34M Ω
Measuring Range
0-33.99M Ω
(Auto-ranging)
Accuracy
± 1.0%rdg ± 3dgt
Buzzer beeps below 30 ± 10Ω
(Continuity buzzer works on 340Ω range
only)
± 5%rdg ± 5dgt
± 15%rdg ± 5dgt
Frequency Hz
Range
Current
Voltage
Measuring Range
0-3.399kHz
3.4kHz-10kHz
(Auto-ranging)
0-3.399kHz
3.4kHz-33.99kHz
34kHz-300kHz
(Auto-ranging)
Accuracy
± 0.1%rdg ± 1dgt
± 0.1%rdg ± 1dgt
●Location for use
Indoor use, Altitude up to 2000m
●Accuracy-insured
23℃±5℃, relative humidity 75% or less
Temperature and
(without condensation)
Humidity Ranges
●Operating Temperature
0-40℃, relative humidity 85% or less
and Humidity Range
(without condensation)
●Storage Temperature
-20-60℃, relative humidity 85% or less
and Humidity Range
(without condensation)
●Source
Two 1.5VDC R03 (UM-4) batteries
●Current Consumption
Approx. 10mA
●Power-save Function
Shifts to the power-save state about 10 minutes after the
last switch operation
(current consumption: approx. 10μA)
●Overload Protection
AC/DC current ranges: MODEL2000 AC/DC 72A for 10
seconds
AC/DC current ranges: MODEL2001 AC/DC
120
A for
10 seconds
AC/DC voltage ranges: AC/DC 720V for 10 seconds
Resistance ranges: AC/DC 720V for 10 seconds
Frequency ranges: AC/DC 720V for 10 seconds
●Withstand Voltage
AC3470V for 5 sec. between electrical circuit and
housing case
●Insulation Resistance
10MΩ or greater at 1000V between electrical circuit and
housing case
●Conductor Size
MODEL2000 Approx. 6mm diameter max.
MODEL2001 Approx. 10mm diameter max.
●Dimensions MODEL2000 128(L)×87(W)×24(D)mm
MODEL2001 128(L)×92(W)×27(D)mm
●Weight
MODEL2000 Approx. 210g
MODEL2001 Approx. 220g
●Accessories Two R03 (UM-4) batteries
Instruction Manual
4.INSTRUMENT LAYOUT
Data Hold Button
MODEL2001 MODEL2000
Range Hold Button
Test Lead
LCD Display
AC/DC Clamp Sensor
DCA
Zero Adjust
Knob
Digital Multimeter
Function Selector Switch
Holster
Barrier
Barrier
AC/DC
Battery Voltage Warning
Auto Range
Data Hold
Unit
Bar Graph
Buzzer
Manual
Range
Wind the lead around the folder.
Test Leads
AC/DC Clamp Sensor
5.PREPARATIONS FOR MEASUREMENT
(1)Checking battery voltage
Set the Function Selector Switch to any position other than the OFF position.
If the marks on the display is clearly legible without symbol
"BATT"
showing, battery
voltage is OK. If the display blanks or
"BATT"
is indicated, replace the batteries
according to section
8: Battery Replacement
.
NOTE
When the instrument is left powered on, the auto-power-save function automatically
shut the power off; The display blanks even if the Function Selector Switch is set to
a position other than the OFF position in this state. To power on the instrument,
turn the Function Selector Switch or press the Data Hold Button. If the display still
blanks, the batteries are exhausted. Replace the batteries.
(2)Make sure that the Function Selector Switch is set to the appropriate range.
Also make sure that data hold function is not enabled. If inappropriate range is
selected, desired measurement cannot be made.
(3)Install Test Lead to the Holster on the side of body
It is possible to measure with seeing the LCD Display keep
Test Lead installing to the Holster.
6.HOW TO MAKE MEASUREMENT
6-1 Current
Measurement
DANGER
● In order to avoid possible shock hazard, never make measurement on circuits with
a maximum voltage difference of 600VAC/DC or greater between conductors
(300VAC/DC or greater between a conductor and ground).
● Do not make measurement with the test leads connected to the circuit under test.
Never make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
CAUTION
● When handling the clamp sensor, exercise caution not to apply excessive shocks
or vibration to the sensor.
●
M
aximum measurable conductor size is MODEL2000 6mm / MODEL2001
10mm in diameter.
AC/DC
Battery Voltage Warning
Auto Range
Data Hold
Unit
Bar Graph
Buzzer
Manual
Range
Wind the lead around the folder.
Test Leads
AC/DC Clamp Sensor
6-1-1 DC Current Measurement
(1)The Function Selector Swith to the " A" position.
("DC" and "AUTO" marks are shown on the top of the display.)
(2) Turn the 0(Zero) ADJ knob to set the reading of the multimeter to zero. (If this zero
adjustment is made incorrectly, measurement errors will result.)
(3) Adjust one of the conductors to the center of the clamp sensor's arrow.
( When the position of the conductor is not at the center of the arrow, the error
occurs.)
Measured value is shown on the display.
Note: When current flows from the upside to the underside of the instrument, the polarity
of the reading is positive (+). Otherwise, the polarity of the reading is negative (-).
6-1-2 AC Current Measurement
(1)Set the Function Selector Switch to " A."
("AC" and "AUTO" marks are shown on the top of the LCD.)
(2) Adjust one of the conductors to the center of the clamp sensor's arrow.
( When the position of the conductor is not at the center of the arrow, the error
occurs.)
Measured value is shown on the display.
Note: Unlike DC current measurement, zero adjustment is not necessary. There is not
polarity indication either.
6-2 Voltage Measurement
DANGER
● In order to avoid possible shock hazard, never make measurement on circuits
with a maximum voltage difference of 600VAC/DC or greater between
conductors (300VAC/DC or greater between a conductor and ground).
● Do not make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
6-2-1 DC Voltage Measurement
(1)Set the Function Selector Switch to " V."
("DC" and "AUTO" marks are shown on the top of the LCD.)
(2) Connect the red test lead to the positive (+) side of the circuit under test and the
black test lead to the negative (-) side. Measured voltage value is shown on the
display.
When the connection is reversed, "-" is shown on the display.
6-2-2 AC Voltage Measurement
(1)Set the Function Selector Switch to " V."
("AC" and "AUTO" marks are shown on the LCD.)
(2) Connect the test leads to the circuit under test.
Measured voltage value is shown on the display.
6-3 Resistance Measurement
DANGER
● Never make measurement on circuits that are live.
● Never make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
Conductor
MODEL2000
MODEL2001
Adjust a
conductor to
the center of
the arrow.
Correct
Wrong
(1)Set the Function Selector Switch to "Ω/ ."
(2) Check that the display shows over-range. Short the test leads and check that the
buzzer beeps and the display reads zero.
(3) Connect the test leads to the circuit under test. Measured resistance value is
shown on the display. When the measured value is below about 30Ω, the buzzer
beeps.
Note: When the test leads are shorted, the display may read a small resistance value.
This is the resistance of the test leads.
If there is an open in either of the test leads, "OL" is shown on the display.
On the 340Ω range, " " is shown on the left side of the LCD.
6-4 Frequency Measurement
DANGER
● In order to avoid possible shock hazard, never make measurement on circuits with
a maximum voltage difference of 600VAC/DC or greater between conductors
(300VAC/DC or greater between a conductor and ground).
● Do not make measurement with the test leads connected to the circuit under test.
Never make measurement with the battery compartment cover removed.
● Do not make current measurement with the test leads connected to the circuit
under test.
●Keep your fingers and hands behind the barrier during measurement.
(1)Set the Function Selector Switch to "Hz."
(2) Measuring frequency of current:
Adjust one of the conductors to the center of the clamp sensor's arrow. Measured
value is shown on the display.
Measuring frequency of voltage:
Connect the test leads to the circuit under test. Measured frequency is shown on
the display.
Note: Measuring range of current frequency is 0-10kHz with minimum measurable input
of MODEL2000 15A(Typ)/MODEL2001 25A(Typ).
Measuring range of voltage frequency is 0-300kHz with minimum measurable
input of 30V(Typ).
When measuring frequency, do not attach the clamp sensor and the test leads to
the circuit under test simultaneously.
7.OTHER FUNCTIONS
7-1 Auto-Power-Save Function
NOTE
A small amount of current is consumed even in the power-save state. Make sure to
set the Function Selector Switch to the OFF position when the instrument is not
used.
Prohibition
This function helps to avoid unwanted exhaustion of the batteries because of
leaving the instrument powered on and extend battery life. The instrument
automatically shifts to the power-save state about 10 minutes after the last
Function Selector Switch or other switch operation.
To return to the normal state: Turn the Function Selector Switch or press the
Data Hold Button twice to exit the power-save state and enable measurement
functions.
7-2 Data Hold Function
This is a function to freeze a measured value on the display. Press the Data
Hold Button once to hold the current reading. In this data hold state, the reading
is held even if input varies. "H" and " " marks are shown on the LCD instead
of "AUTO" mark.
To exit the data hold state, press the Data Hold Button again.
7-3 Range Hold Function
The instrument defaults to auto-ranging ("AUTO" is shown on the LCD).
Pressing the Range Hold Button enables manual selection among measurement
ranges(" " mark is shown on the LCD instead of "AUTO" mark)
Press the Range Hold Button to select a higher range.
To switch from manual range selection to auto-ranging, press down the Range
Hold Button for about one seconds, or turn the Function Selector Switch to
another position before setting it back to the current range.
8.BATTERY REPLACEMENT
WARNING
● In order to avoid possible shock hazard, always disconnect the test leads from
the circuit under test and set the Function Selector Switch to the OFF position
before trying to replace the batteries.
CAUTION
● Do not mix new and old batteries.
● Install batteries in the orientation as shown inside the battery compartment,
observing correct polarity.
When the battery voltage warning mark "BATT" is shown on the top left corner of the
LCD, replace the batteries. Note that the display blanks and "BATT" mark is not
shown if the batteries are completely exhausted.
(1) Set the Function Selector Switch to "OFF."
(2) Remove the instrument from the holster.
(3) Loosen the battery-compartment-cover-fixing screw on the lower back of the
instrument.
(4) Replace the batteries with two new R03 (UM-4) 1.5V batteries.
(5) Put the battery compartment cover back in place and tighten the screw.
Screw
Battery
Compartment
Cover
Batteries
92-1507E
8-16
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or
designs described in this manual without notice and without
obligations.