Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-290
3700AS-901-01 Rev. D/June 2018
A local node reset includes a
channel.reset("allslots")
,
dmm.reset("all")
and a
scan.reset()
. In addition:
•
Other settings are restored to factory default settings
•
Existing channel patterns and DMM configurations are deleted
•
All channels and backplane relays are opened
•
The dmm function is
"dcvolts"
•
User-created reading buffers are deleted
A
localnode.reset()
is different than a
reset()
because
reset()
resets the entire system.
When using this command from a remote node,
localnode
should be replaced with the node
reference, for example
node[5].reset()
.
Example
localnode.reset()
Resets the local node.
Also see
(on page 11-84)
(on page 11-219)
(on page 11-308)
(on page 11-324)
localnode.revision
This attribute stores the firmware revision level.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
revision
= localnode.revision
revision
Firmware revision level
Details
This attribute indicates the revision number of the firmware that is presently running in the instrument.
When using this command from a remote node,
localnode
should be replaced with the node
reference. For example,
node[5].revision
.
Example
print(localnode.revision)
Outputs the present revision level.
Sample output:
01.50b
Also see
(on page 11-284)
(on page 11-291)