Instruction
Manual
F
o
r
P
ro
fe
ss
io
n
a
ls
..
.S
in
ce
1
8
5
7
®
ENGLISH
CL1000
V
AT
0
10
20
40
30
Apo
600V
400A
CATIII
CL1000
• DATA HOLD
• BACKLIGHT
• WORKLIGHT
• BAR GRAPH
• MAX
/
MIN
• 3-3/4 DIGIT
3999 COUNT LCD
• AUTO / MANUAL
RANGE
• NON CONTACT
VOLTAGE TESTER
SyMBOLS USED ON LCD
~
AC Measurement
DC Measurement
-
Negative DC Value
AT
Auto Range Active
O.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
MAX
Maximum Reading
%
Duty Cycle Mode
Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms
Diode Test
F
Capacitance in Farads
Continuity Test
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
PRESS
and
HOLD
600V
400V
CA
TII
I
CL1000
PRESS
and
HOLD
8.
Non Contact Voltage (NCV): > 25V AC
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range
Resolution
Accuracy
400mV ˜ 400V
0.1mV ˜ 0.1V
± (0.5% + 4 digits)
1000V
1V
± (0.8% + 10 digits)
Overload Protection:
1000V
AC Voltage Measurement
Range
Resolution
Accuracy
400mV ˜ 750V
0.1mV ˜ 1V
± (2.0% + 5 digits)
Overload Protection:
750V RMS
Frequency Response:
40 to 400 Hz
Minimum Voltage for Frequency Measurement:
200mV
Response:
Averaging
DC Current Measurement
Range
Resolution
Accuracy
400µA
0.1µA
± (1.2% + 3 digits)
2000µA
1µA
Overload Protection:
•
Voltage:
600V RMS
•
Current:
2000µA
AC Current Measurement
Range
Resolution
Accuracy
400µA
0.1µA
± (2.0% + 5 digits)
2000µA
1µA
± (1.5% + 5 digits)
40A
0.01A
± (2.9% + 15 digits)
400A
0.1A
± (1.9% + 8 digits)
Overload Protection:
•
Voltage:
600V RMS
•
Current:
2000µA
Frequency:
45 to 400Hz
Minimum Current for Frequency Measurement:
400µA or 20A
Response:
Averaging
Resistance Measurement
Range
Resolution
Accuracy
400
Ω
˜ 4M
Ω
0.1
Ω
˜ 0.001M
Ω
± (1.0% + 4 digits)
40M
Ω
0.01M
Ω
± (2.0% + 4 digits)
Overload Protection:
600V RMS
Capacitance Measurement
Range
Resolution
Accuracy
40nF ˜ 4000µF
0.01nF ˜ 1µF
± (3.5% + 6 digits)
Overload Protection:
600V RMS
Frequency Measurement
Range
Resolution
Accuracy
9.999Hz ˜ 999.9kHz
0.001Hz ˜ 0.1kHz
± (0.1% + 4 digits)
Overload Protection:
600V RMS
Duty Cycle Measurement
Range
Resolution
Accuracy
0.1 ˜ 99.9%
0.1%
± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection:
600V RMS
Diode Test
Overload Protection Range Test Current
Open Circuit Voltage
600V RMS
2.0V
Appx. 0.25mA
< 1.6V DC
Continuity Test
Overload Protection
Open Circuit Voltage
600V RMS
< 0.44V
Non Contact Voltage Detector
On Voltage
Appx. 25V AC
WARRANTy
This product is warranted to be free from defects in materials and workmanship
for a period of two years from the date of purchase. During this warranty period,
Klein Tools has the option to repair or replace or refund the purchase price of
any unit which fails to conform to this warranty under normal use and service.
This warranty does not cover damage which occurs in shipment or failure which
results from alteration, tampering, accident, misuse, abuse, neglect, or improper
maintenance. Batteries and damage resulting from failed batteries are not
covered by warranty. A purchase receipt or other proof of original purchase date
will be required before warranty repairs will be rendered.
Any implied warranties, including but not limited to implied warranties of
merchantability and fitness for a particular purpose, are limited to the express
warranty. Klein Tools shall not be liable for loss of use of the instrument or other
incidental or consequential damages, expenses, or economic loss, or for any
claim or claims for such damage, expenses or economic loss.
Some states or countries laws vary, so the above limitations or exclusions may
not apply to you. This warranty gives you specific legal rights, and you may also
have other rights which vary from state to state. If your Klein product requires
repair or for information on how to exercise your rights under the terms of this
warranty, please contact Klein Tools at 1-877-775-5346.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECyCLE
Caution:
This symbol indicates that equipment and its accessories
shall be subject to a separate collection and correct disposal.