Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-335
Trigger event IDs
Trigger event ID
Description
scan.trigger.EVENT_SCAN_READY or 24
Scan ready event.
scan.trigger.EVENT_SCAN_START or 25
Scan start event.
scan.trigger.EVENT_CHANNEL_READY or 28
Channel ready event.
scan.trigger.EVENT_MEASURE_COMP or 56
Measure complete event.
scan.trigger.EVENT_SEQUENCE_COMP or 50
Sequence complete event.
scan.trigger.EVENT_SCAN_COMP or 26
Scan complete event.
scan.trigger.EVENT_IDLE or 27
Idle event.
schedule.alarm[N].EVENT_ID or 37 to 38
Trigger event generated by the alarm
N
.
Use one of the text trigger event IDs (for example,
digio.trigger[
N
].EVENT_ID)
to set the
stimulus value rather than the numeric value. Doing this will make the code compatible for future
upgrades.
Example
scan.trigger.measure.stimulus = scan.trigger.EVENT_CHANNEL_READY
Sets the trigger stimulus of the measurement event detector to the channel ready event.
Also see
scan.trigger.sequence.stimulus
(on page 11-336)
(on page 5-23)
scan.trigger.sequence.clear()
This function clears the sequence event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
scan.trigger.sequence.clear()
Details
This function sets the sequence event detector to the undetected state.
Example
scan.trigger.sequence.clear()
Clears the sequence event detector.
Also see
(on page 5-23)