Model 2700 Multimeter/Switch System User’s Manual
Scanning
7-25
2.
Press the
Δ
or
∇
key to display IMM SCAN: N and press ENTER.
a.
Press the
Δ
or
∇
key to enable or disable low limit 1 (LLIM1 SCAN:N/Y),
and press ENTER.
b.
Press the
Δ
or
∇
key to enable or disable high limit 1 (HLIM1 SCAN:N/Y),
and press ENTER.
c.
Press the
Δ
or
∇
key to enable or disable low limit 2 (LLIM2 SCAN:N/Y),
and press ENTER.
d.
Press the
Δ
or
∇
key to enable or disable high limit 2 (HLIM2 SCAN:N/Y),
and press ENTER.
3.
Finish configuring the scan by performing Step 4 and Step 5 of the
.
NOTE
For a remote programmed monitor scan, use the ROUTe:MONitor:POINts
command to specify the number of channels to scan (
).
4.
While in the normal measurement state, select and enable the monitor channel as
explained in “
—
.
When the reading limit for the monitor channel is reached, the scan will be
triggered to start. When the monitor channel is scanned, the display will show the
reading that triggered the scan.
If the reading limit event is still present on the monitor channel when the scan
finishes, the scan will be triggered to run again. Note that the scan can also be run
by pressing STEP or SCAN.
5.
To disable the monitor scan, perform the following steps:
a.
To disable monitor, press SHIFT and then MONITOR (MON annunciator
turns off).
b.
To disable limits, press SHIFT and then ON/OFF. Press
Δ
or
∇
to display
“LIMITS: OFF” and press ENTER.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......