Model 2700 Multimeter/Switch System User’s Manual
FORMat and Misc SYSTem Commands
14-7
Limits
— For the ASCII data format, limit test results are returned as a 4-bit binary
number “abcd” where:
a = High limit 2
b = Low limit 2
c = High limit 1
d = Low limit 1
A “0” indicates that the limit has passed, while a “1” indicates that the limit has failed.
For the binary data formats, the limits information must be decoded from the returned
value (0 to 15). Convert the value to its binary equivalent for “abcd” where “d” is the LSD
and “a” is the MSD. For example, the value 10 converted to its binary equivalent is1010.
That means High Limit 2 and High Limit 1 have failed.
FORMat:BORDer <name>
Parameters
NORMal
= Normal byte order for IEEE-754 binary format
SWAPped
= Reverse byte order for IEEE-754 binary format
For normal byte order, the data format for each element is sent as follows:
Byte 1
Byte 2
Byte 3
Byte 4
(Single precision)
Byte 1
Byte 2
...
Byte 8
(Double precision)
For reverse byte order, data is sent as follows:
Byte 4
Byte 3
Byte 2
Byte 1
(Single precision)
Byte 8
Byte 7
...
Byte 1
(Double precision)
) is not affected by this command. The header is always sent
at the beginning of the data string.
The ASCII data format can only be sent in the normal byte order. The SWAPped selection
is ignored when the ASCII format is selected.
NOTE
The SWAPped byte order must be used when transmitting binary data to any
IBM PC.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......