Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-59
[:SENSe[1]]:DATA[:LATest]?
What it does
This query will return the last reading the instrument had, regardless of what may have
invalidated that reading, such as changing ranges or functions.
Limitations
This query is fully capable of returning meaningless, old data.
When appropriate
If, for some reason, the user wanted the last completed reading, even after changing ranges
or other measurement settings, which would invalidate the old reading.
The :CALC:DATA:LATest? query is similar to the :DATA:LAT? query, but applies to
readings that have math applied to them (e.g.,:MX+B scaling).
Examples
One-shot reading, DC volts, no trigger, fastest rate
*RST
:INITiate:CONTinuous OFF;:ABORt
:SENSe:FUNCtion ‘VOLTage:DC’
:SENSe:VOLTage:DC:RANGe 10
// Use fixed range for fastest readings.
:SENSe:VOLTage:DC:NPLC 0.01
// Use lowest NPLC setting for fastest
// readings
.
:DISPlay:ENABle OFF
// Turn off display to increase speed.
:SYSTem:AZERo:STATe OFF
// Disable autozero to increase speed, but
// may cause drift over time
.
:SENSe:VOLTage:DC:AVERage:STATe OFF
// Turn off filter for
speed.
:TRIGger:COUNt 1
:READ?
(Enter reading)
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......