7-20
Scanning
Model 2700 Multimeter/Switch System User’s Manual
Auto channel configuration
Auto channel configuration allows you to recall scan list setups. With auto channel
configuration enabled, a closed channel assumes the scan list setup. With this feature, you
can inspect the channel setups of the scan, or manually scan channels. When a scan
channel is disabled (not in scan list), it cannot be closed with auto channel configuration
enabled.
As with normal operation, when you use the
,
, or CLOSE to close a channel (or
channel pair), any other closed channels are first opened.
Perform the following steps to enable or disable auto channel configuration.
1.
Press SHIFT and then SETUP.
2.
Use the
Δ
or
∇
key to display the auto configuration (CH AUTOCFG) setting;
N (no) or Y (yes).
3.
Press the
key to place the cursor on the present setting (N or Y), and press the
Δ
or
∇
key to change the setting.
4.
Press ENTER to return to the normal measurement state.
NOTES
Auto channel configuration cannot be enabled if there is a non-scan channel
presently closed. For example, assume the scan list consists of channels 105
through 110, and channel 101 (a non-scan channel) is presently closed. When
you attempt to enable auto channel configuration from the front panel, the
message “NOT IN SCAN” is briefly displayed. For remote operation, error -221
(settings conflict) occurs.
With auto channel configuration enabled, the
and
, keys will not properly
step through a non-sequential scan list. Therefore, auto channel configuration
should not be used for a non-sequential scan list. For information on non-
sequential scans, see “
. This topic is located near the beginning of this section.
For remote operation, the ROUT:CLOS:ACON command is used to enable or
disable auto channel configuration (
).
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......