Model 2700 Multimeter/Switch System User’s Manual
Measurement Considerations
E-9
Meter loading
Loading of the voltage source by the Model 2700 becomes a consideration for high source
resistance values. As the source resistance increases, the error caused by meter loading
increases.
shows the method used to determine the percent error due to meter loading.
The voltage source, V
S
, has a source resistance, R
S
, while the input resistance of the
Model 2700 is R
I
, and the voltage measured by the nanovoltmeter is V
M
.
Figure E-5
Meter loading
The voltage actually measured by the meter is attenuated by the voltage divider action of
R
S
and R
I
, and it can be calculated as follows:
This relationship can be modified to directly compute for percent error:
From the above equation, it is obvious that the input resistance of the Model 2700 must be
at least 999 times the value of source resistance if loading error is to be kept to within
0.1%.
V
M
Measured
Voltage
R
S
R
I
V
S
Source
Voltage
+
–
V
M
V
S
R
L
R
I
R
S
+
-------------------
=
Percent error
100R
S
R
I
R
S
+
-------------------
=
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......