Model 2700 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-17
Basic operation
NOTE
Make sure the INPUTS switch is set to the REAR position (in).
1.
Select and configure (range, filter, rel, etc.) a valid measurement function. For
ratio, the only valid function is DCV. For channel average, the only valid functions
are DCV and TEMP (TCs only).
2.
Use the
or
key to select (close) a primary channel (101 through 110 for the
Model 7700). The CLOSE key can also be used.
3.
Apply one signal to the selected primary channel, and apply the other signal to the
paired channel. For the Model 7700, if the closed primary channel is 101, the
paired channel is 111.
4.
Enable Ratio or Channel Average:
•
Ratio — Press SHIFT and then RATIO. The RATIO annunciator will turn on
to indicate that the displayed readings are the result of the ratio calculation.
•
Channel Average — Press SHIFT and then CHA-AVG. The DELTA
annunciator will turn on to indicate that the displayed readings are the result of
the channel average calculation. To disable channel average, again press
SHIFT and then CH-AVG.
5.
When finished, there are two ways to disable the calculation:
•
Press the OPEN key. The calculation will disable and the channel will open.
•
Press SHIFT and then RATIO to disable ratio, or press SHIFT and then
CHA-AVG to disable channel average. The calculation will disable, but the
channel will remain closed.
NOTE
The paired channel number is not displayed when it is measured. Only the
primary channel is displayed during the 2-channel scan for the calculation.
Enabling ratio disables channel average and conversely, enabling channel
average disables ratio.
If either of the channel readings over range (“OVRFLW”), the result of the
calculation will also be “OVRFLW.”
When using limits with ratio or channel average, the limit values will be
compared to the result of the calculation and not to the individual channels.
With ratio or channel average enabled, pressing a function key will display the
“EXIT RATIO” or “EXIT CHA-AVG” message to indicate that the calculation
must first be disabled as explained in step 5 of the above procedure.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......