3-6
Basic DMM Operation
Model 2700 Multimeter/Switch System User’s Manual
Remote programming — autozero and LSYNC
Autozero and LSYNC commands
The commands to control display resolution (digits) are listed in
Channel list parameter (<clist>)
Channels of one or more switching modules installed in the Model 2700 can be scanned.
Each scan channel can have its own unique setup. For example, a channel could be set to
measure DCV on the 10V range, while another channel can be set to measure ACV on the
1V range.
From the front panel, scan channels are configured from the scan configuration menu as
explained in
. For remote programming, the <clist> parameter is used to
configure scan channels.
Channel list parameter:
<clist> = (@SCH)
where:
S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Table 3-1
Autozero and LSYNC commands
Commands
Description
Default
Autozero command*
SYSTem:AZERo[:STATe] <b>
Enable or disable autozero; <b> = ON or
OFF
ON
Line synchronization command
SYSTem:LSYNc[:STATe] <b>
Enable or disable LSYNC; <b> = ON or
OFF
OFF
* After enabling autozero, you can update the internal reference points immediately by setting the integration
rate to 0.01 PLC and then back to the desired setting (see NPLC commands in
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......