3-18
Basic DMM Operation
Model 2700 Multimeter/Switch System User’s Manual
Model 7700 switching module
Connections for the Model 7700 switching module are shown in
. Note that
only channels 21 and 22 can be used for current measurements.
Figure 3-7
DCI and ACI connections using Model 7700 switching module
Amps measurement procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it has to be in the “F” (out) position. For switching modules, it has to be
in the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the amps measurement function by pressing DCI or ACI.
3.
Use the RANGE
Δ
and
∇
keys to select a measurement range consistent with the
expected current, or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in
.
4.
Apply the current(s) to be measured.
CAUTION
Do not apply more than 3A to the input or the AMPS fuse will blow.
Model 7700 switching module — When performing current
measurements, the maximum allowable voltage is 60VDC or 30V
RMS. Exceeding these limits could cause damage to the switching
module.
5.
If using a switching module, use the
and
keys to close the desired amps
channel (for the Model 7700, 21 or 22). All other channels will be open.
Model 7700
Switching
Module
H
L
CH 21 or 22
Caution:
Maximum input: 60VDC or 30V RMS, 3A switched,
60W, 125VA maximum
Current
Source
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......