12-4
Common Commands
Model 2700 Multimeter/Switch System User’s Manual
After addressing the Model 2700 to talk, the returned value of 1 denotes that the bit (bit 1)
is set indicating that the :INITiate operation is now complete.
SYST:PRES
‘ Returns 2700 to default setup.
NOTE
The following commands take a long time to process and may benefit from
using *OPC or *OPC?:
•
*RST and SYST:PRES
•
*RCL and *SAV
•
ROUT:MULT:CLOS and ROUT:MULT:OPEN – Only if the <clist> is long.
•
CALC2:IMM and CALC2:IMM?– Only when performing the standard devi-
ation calculation on a large buffer. A 10,000 point buffer takes around 5.75
seconds.
C *OPC?
⎯
Operation Complete Query
Place a ‘1’ in the output queue after
all pending operations are completed
Description
When this common command is sent, an ASCII “1” will be placed in the Output Queue
after the last pending operation is completed. When the Model 2700 is then addressed to
talk, the “1” in the Output Queue will be sent to the computer.
The “1” in the Output Queue will set the MAV (Message Available) bit (B4) of the Status
Byte Register. If the corresponding bit (B4) in the Service Request Enable Register is set,
the RQS/MSS (Request for Service/Master Summary Status) bit in the Status Byte
Register will set.
When used with the Initiate Immediately command (:INITiate), a “1” will not be placed
into the Output Queue until the Model 2700 goes back into the idle state. The :INIT
command operation is not considered finished until the Model 2700 goes back into the idle
state. See the description for *WAI for more information on command execution.
The execution of OPC? is not completed until it has placed the “1” in the Output Queue.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......