Model 2700 Multimeter/Switch System User’s Manual
Scanning
7-9
The auto delay period cannot be adjusted by the user. It is a fixed delay for the selected
function and range (
NOTE
When scanning, the auto delay times in
are valid for all control
sources (Immediate, External, Timer, Manual, or Bus).
With manual delay selected, the user can set the delay period from 0 seconds to 99 hours,
99 minutes, 99.999 seconds. However, if you set a delay shorter than the corresponding
auto delay period, measurement uncertainty increases (noisy and/or unsettled readings
may result).
NOTE
Keep in mind that if the timer control source is selected, the Delay period is only
in effect for the first pass through the loop.
Ratio/Chan Average Delay
— With ratio or channel average enabled, a delay is typically
used to keep the channel relays from cycling too fast. The default delay period is
0.5 seconds but can be set from 0 to 999999.999 seconds using remote programming.
Ratio and channel average are covered in
.
NOTE
The Ratio/Chan Average Delay is in addition to the Timer or Delay (Auto or
Manual). That is, it occurs after the Timer interval or Delay period elapses.
Device action
The channel measurement process is performed at this block. If repeat filter is enabled, the
filter process is also performed.
Reading count
NOTE
For both STEP and SCAN, the reading count specifies the number of readings to
store in the buffer.
STEP operation
— The reading count specifies the number of channels to scan. This can
be equal to, less than, or greater than the number of channels in the scan list. The last
scanned channel remains closed. If you start the scan again, it will start at the next
channel.
If the reading count is set to infinity (INF), the scan will continuously repeat until you
stop it.
NOTE
One counter is used for STEP operation. As shown in
, reading count
sets the Trigger Counter.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......