4-6
Range, Digits, Rate, Bandwidth, and Filter
Model 2700 Multimeter/Switch System User’s Manual
Scanning
When a simple scan is configured, the present digits setting will apply to all channels in
the scan. When an advanced scan is configured, each channel can have its own unique
digits setting. Details to configure and run a scan are provided in
.
For remote programming, the <clist> parameter is used to configure channels for a scan.
Remote programming — digits
Digits commands
The commands to control display resolution (digits) are listed in
. Additional
information on these commands follow the table.
NOTE
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Table 4-3
Digits commands
Commands*
Description
Default
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:DIGits <n> [, <clist>]
Set # of digits for DCV; <n> = 4 to 7.
7
:VOLTage:AC:DIGits <n> [, <clist>]
Set # of digits for ACV; <n> = 4 to 7.
6
:CURRent[:DC]:DIGits <n> [, <clist>]
Set # of digits for DC1; <n> = 4 to 7.
7
:CURRent:AC:DIGits <n> [, <clist>]
Set # of digits for ACI; <n> = 4 to 7.
6
:RESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
2; <n> = 4 to 7.
7
:FRESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
4; <n> = 4 to 7.
7
:TEMPerature:DIGits <n> [, <clist>]
Set # of digits for TEMP; <n> = 4 to 7.
6
:FREQuency:DIGits <n> [, <clist>]
Set # of digits for FREQ; <n> = 4 to 7.
7
:PERiod:DIGits <n> [, <clist>]
Set # of digits for PERIOD; <n> = 4 to 7.
7
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
* The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the digits command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:DIGits 4.5, (@101) is only
valid if scan channel 101 is set for the ACV function.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......