5-4
Rel, Math, Ratio, Channel Average, dB
Model 2700 Multimeter/Switch System User’s Manual
Remote programming — rel
Rel commands
The rel commands to set range are listed in
. Additional information on these
commands follow the table.
NOTE
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Table 5-1
Rel commands
Commands
1
Description
Default
Rel commands for DCV
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:REFerence <n> [, <clist>]
Specify rel value; <n> = -1010 to 1010
(V).
0
:VOLTage[:DC]:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:VOLTage[:DC]:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for ACV
[SENSe[1]]
Optional root command.
:VOLTage:AC:REFerence <n> [, <clist>]
Specify rel value; <n> = -757.5 to 757.5
(V).
0
:VOLTage:AC:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:VOLTage:AC:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for DCI
[SENSe[1]]
Optional root command.
:CURRent[:DC]:REFerence <n> [, <clist>]
Specify rel value; <n> = -3.1 to 3.1 (A).
0
:CURRent[:DC]:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:CURRent[:DC]:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for ACI
[SENSe[1]]
Optional root command.
:CURRent:AC:REFerence <n> [, <clist>]
Specify rel value; <n> = -3.1 to 3.1 (A).
0
:CURRent:AC:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:CURRent:AC:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......