7-36
Scanning
Model 2700 Multimeter/Switch System User’s Manual
Monitor scan
For this example, channel 101 of the Model 7700 is used to monitor temperature. When
the temperature reading reaches 30°C, it will start the scan. For this 4-channel scan,
channel 101 measures temperature, while channels 102, 103, and 104 measure DCV.
This example uses the channel average feature to measure temperature. With channel
average enabled, two temperature measurements will be taken; one at channel 101 and
another at its paired channel (111). The two measured readings are then averaged to yield
a single reading. It is this averaged temperature reading that will start the scan when it
reaches 30°C. Refer to
for details on channel average.
Two type K thermocouples are used to measure temperature. Since the internal cold
(reference) junction of the Model 7700 is being used, the thermocouples can be connected
directly to the screw terminals of the switching module.
Operation
A simplified model of monitor scan operation is shown in
steps and programming commands are listed in
, notice that there are two modes of operation. While in the monitor mode,
the Model 2700 continuously performs temperature measurements. Keep in mind that
channel average is being used. Therefore each temperature reading is the average of two
temperature measurements (one on channel 101 and one on channel 111). As long as the
average temperature reading remains below 30°C, the instrument will remain in the
monitor mode.
When the temperature reading reaches 30°C, the Model 2700 switches over to the scan
mode. Figuratively speaking, it is as if a “finger” presses the SCAN key when the monitor
detects that the average temperature is at or above 30°C.
The instrument is configured to scan four channels. The monitor TEMP channel reading
and three DCV channel readings are stored in the buffer. After the fourth channel is
measured, operation returns to the monitor mode, to again measure temperature. Note that
if the average temperature is still at or above 30°C, the “finger” will again press SCAN to
start the scan.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......