Model 2700 Multimeter/Switch System User’s Manual
SCPI Reference Tables
15-19
[SENSe[1]]
:CONTinuity
Path to configure continuity test:
:THReshold <NRf>
Set threshold resistance in ohms
(1 to 1000).
10
:THReshold?
Query threshold resistance.
Notes:
1.
CAVerage:DELay and RATio:DELay are coupled. Changing the delay for channel average also changes the delay for channel
ratio, and vice versa.
2.
For 60Hz line power, the default for aperture is 16.67msec. For 50Hz, the default is 20msec.
3.
REPeat is the *RST default and MOVing is the SYSTem:PRESet default. From the front panel, the factory default is MOVing.
4.
OFF is the *RST default, and ON is the SYTem:PRESet default.
5.
The following commands can instead be used to select the reference junction and set the simulated reference temperature:
TEMPerature:RJUNction:RSELect <name> [, <clist>]
Select reference junction; SIMulated, INTernal, or EXTernal.
TEMPerature:RJUNction:RSELect? [<clist>]
Query reference junction.
TEMPerature:RJUNction:SIMulated <n> [, <clist>]
Set simulated reference temperature; 0 to 50 (°C), 32 to 122 (°F),
or 273 to 323 (K).
TEMPerature:RJUNction:SIMulated? [<clist>]
Query simulated reference temperature.
6.
When using multiple channel operation (ROUT:MULT command) to connect a switching module channel to the DMM for ther-
mocouple temperature measurements, the SIMulated reference junction will be used if the INTernal or EXTernal reference junc-
tion is selected.
7. With a Model 7700, 7706, or 7708 installed, the default sensor junction is Internal. Otherwise, the Simulated (23°C) junction is
selected.
8. Only one USER RTD per scan list.
Table 15-5 (continued)
SENSe command summary
Command
Description
Default
parameter
Ref
SCPI
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......