5-6
Rel, Math, Ratio, Channel Average, dB
Model 2700 Multimeter/Switch System User’s Manual
“Pressing REL” using rel commands
When the front panel REL key is pressed, the displayed reading is used as the rel value.
Subsequent readings are then the result of the actual input value and the rel value.
The :REFerence:ACQuire and :REFerence:STATe ON commands (in that order) can be
used to “press” the REL key. For example, the following command sequence is the
equivalent of pressing the REL key while on the DCV function:
VOLT:REF:ACQ
' Acquire reading as rel value.
VOLT:REF:STAT ON
' Enable rel.
Setting rel values
The :REFerence <n> command specifies the rel value for the specified function, while the
:ACQuire command uses the input signal as the rel value. The :ACQuire command is
typically used to zero the display. For example, if the instrument is displaying a 1µV
offset, sending :ACQuire and enabling rel (STATe ON) zeroes the display.
The ACQuire command is only functional if the instrument is on the specified function.
For example. If the instrument is on the DCV function, the only valid acquire command is
VOLT:DC:REF:ACQ. Also, if the instrument is overflowed (“OVERFLOW”), or a read-
ing has not been triggered (“------”), an execution error (-200) occurs when :ACQuire is
sent.
The :REFerence <n> command is coupled to the ACQuire command. When a rel value is
set using :REFerence <n>, the :REFerence? query command returns the programmed
value. When rel is set using :ACQuire, the :REFerence? query command returns the
acquired rel value.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......