13-4
SCPI Signal Oriented Commands
Model 2700 Multimeter/Switch System User’s Manual
CONFigure:<function> [<rang>], [<res>], [<clist>]
CONFigure:VOLTage[:DC] [<rang>], [<res>], [<clist>]
Configure DCV
CONFigure:VOLTage:AC [<rang>], [<res>], [<clist>]
Configure ACV
CONFigure:CURRent[:DC] [<rang>], [<res>], [<clist>]
Configure DCI
CONFigure:CURRent:AC [<rang>], [<res>], [<clist>]
Configure ACI
CONFigure:RESistance [<rang>], [<res>], [<clist>]
Configure
Ω
2
CONFigure:FRESistance [<rang>], [<res>], [<clist>]
Configure
Ω
4
CONFigure:FREQuency [<rang>], [<res>], [<clist>]
Configure FREQ
CONFigure:PERiod [<rang>], [<res>], [<clist>]
Configure PERIOD
CONFigure:TEMPerature [<clist>]
Configure TEMP
CONFigure:CONTinuity [<clist>]
Configure CONT
Parameters
<rang> = Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See
for additional information.
<res> =
0.1
i.e., 100.0 V (3
H
digits)
0.01
i.e., 10.00 V (3
H
digits)
0.001
i.e., 1.000 V (3
H
digits)
0.0001
i.e., 1.0000 V (4
H
digits)
0.00001
i.e., 1.00000 V (5
H
digits)
0.000001
i.e., 1.000000 V (6
H
digits)
The resolution of the <res> parameter value and the selected range
sets the number of display digits. As shown above, with the 100V
range selected and <res> = 0.1, a 100V reading will be displayed
as 100.0 V (3
H
digits).
The display will default to 3
H
digits when using parameter values
that attempt to set the display below 3
H
digits. For example, a 10V
reading using <res> = 0.1 for the 10V range is displayed as 10.00
V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7
H
digits is ignored, and generates error -221
(settings conflict error).
The <res> parameter is ignored when a <clist> is included in
the command string. Resolution for the scanlist channel(s) is
determined by the present setting for the specified function and
by the present resolution setting for the specified function.
for additional
information.
<clist> =
Channel(s) in the scanlist to be configured. When the channel
list parameter (<clist>) is included, the present instrument
settings are not affected. Instead, the channel(s) in the <clist>
for the specified function is configured. See the “NOTES” that
follow
for additional information.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......