13-8
SCPI Signal Oriented Commands
Model 2700 Multimeter/Switch System User’s Manual
MEASure:<function>? [<rang>], [<res>], [<clist>]
MEASure:VOLTage[:DC]? [<rang>], [<res>], [<clist>]
Measure DCV
MEASure:VOLTage:AC? [<rang>], [<res>], [<clist>]
Measure ACV
MEASure:CURRent[:DC]? [<rang>], [<res>], [<clist>]
Measure DCI
MEASure:CURRent:AC? [<rang>], [<res>], [<clist>]
Measure ACI
MEASure:RESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
2
MEASure:FRESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
4
MEASure:FREQuency? [<rang>], [<res>], [<clist>]
Measure FREQ
MEASure:PERiod? [<rang>], [<res>], [<clist>]
Measure PERIOD
MEASure:TEMPerature? [<clist>]
Measure TEMP
MEASure:CONTinuity? [<clist>]
Measure CONT
Parameters
<rang>
= Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See the
“NOTES” that follow
for additional information.
<res>
= 0.1
i.e., 100.0 V (3
H
digits)
0.01
i.e., 10.00 V (3
H
digits)
0.001
i.e., 1.000 V (3
H
digits)
0.0001
i.e., 1.0000 V (4
H
digits)
0.00001
i.e., 1.00000 V (5
H
digits)
0.000001
i.e., 1.000000 V (6
H
digits)
The resolution of the <res> parameter value and the selected
range sets the number of display digits. As shown above, with
the 100V range selected and <res> = 0.1, a 100V reading will
be displayed as 100.0 V (3
H
digits).
The display will default to 3
H
digits when using parameter
values that attempt to set the display below 3
H
digits. For
example, a 10V reading using <res> = 0.1 for the 10V range is
displayed as 10.00 V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7
H
digits is ignored, and generates error -221
(settings conflict error).
for additional
information.
<clist>
= Single channel only. When included, this is the channel to be
closed and measured.
Description
The MEASure? command combines all of the other signal oriented
measurement commands to perform a “one-shot” measurement and
acquire the reading. If the <clist> parameter is included, the specified
channel will close before performing the measurement.
When a MEASure? command is sent, the specified function is selected.
If specified, range and resolution will also set.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......